Analysis system | |
Working mode | Discrete/random access |
Test speed | test speed 300T/H (without ISE) |
Test principle | Absorbance photometry, Turbidimetry |
Methodology | End-point, Fixed-time, Kinetic, Single/Dual reagent chemistries, monochromatic/bichromatic Linear/non-linear multi-point calibration |
The longest reaction time | 15 minutes |
Minimum reaction volume | 150μl |
Cuvettes material | plastics (quartz glass can be selected), 81 cuvettes have the function of automatically deducting reagent and sample blanks |
Linear range of absorbance | 0-3.8Abs |
Halogen light source using time | ≥ 2000 hours |
Wavelength | 340nm、405nm、450nm、510nm、546nm、578nm、630nm、670nm |
Refrigeration system | water medium uniform refrigeration technology |
Reagent storehouse temperature | 4-16℃ |
Reaction temperature | 37±0.2℃ |
Temperature fluctuation | ±0.1℃ |
Sampling and mixing system | |
Sample capacity | 2-30μl, 0.1μl increasing |
Sample probe | The inner and outer walls are highly polished,with the function of liquid level detection, tracking and collision protection,sample probe matching special cleaning liquid,prevent cross contamination |
Reagent capacity | 2-300μl, 0.1μl increasing |
Reagent probe | The inner and outer walls are highly polished,with the function of liquid level detection, tracking and collision protection |
Reagent bottle | Reagent bottles with volume of 15ml, 20ml, 30ml, 50ml, 60ml and 70ml can be used, and the dead volume is less than 1ml |
Mixing bar | Double mixing bar, S and R2 independent mixing bar to reduce cross contamination. The mixing bar is made of special material, surface treatment using Teflon, not hanging liquid, thoroughly clean before and after mixing to prevent cross contamination |
Reagent Handing | |
Reagent tray | 80 positions in refrigerated compartment(4~16℃) |
Reagent volume | R1:150~300 μL, R2:20~150 μL |
Reagent probe | Liquid level detection, collision protection and inventory check |
Probe cleaning | Interior and exterior automatic probe washing |
Sample Handing | |
Sample tray | 120 positions for sample primary or secondary tubes and sample cups |
Sample volume | 2~100μl, step by 0.1μl |
Sample probe | Liquid level detection, clot detection and collision protection |
Probe cleaning | interior and exterior automatic probe washing carry-over<0.05% |
Calibration and quality control | |
Quality control type | real time quality control, daily quality control, day to day quality control |
Quality control charts | L-J, Cumulative, Twin Plot |
The quality control rules | arbitrarily formulated by default to Westguard multiple rules |
The calibration type | linear and nonlinear. Logit-4P, Logit-5P, Spline, exponent, polynomial, factor method |
Calibration | Automatically check the calibration curve and select the best calibration type of the fitting degree automatically |
Operating System | |
English operating system: | Windows2000、windows XP、windows7、windows8、windows10 |
Report printing | Make print format and content freely, provide several common formats |
user management | Specialmanagement password, rational distribution ofoperation permissions |
Interface | RS-232 standard interface, support for intranet, remote print report |
Databases | Optional various databases, autosave, automatic backup, save all kinds of data permanently |
Real-time monitoring of sample tray、reagents tray、reacting tray;Real-time display of reaction temperature, reagent allowance, reaction curve, calibration curve and quality control chart | |
Check and judge reaction endpoint, reaction linear interval, substrate exhaustion, reagent blank absorbance, etc. | |
Language | English, Chinese |
其它相關店家資訊